Michelle Sensitivity


The values below are based on calculations with the Michelle Integration Time Calculator (ITC). The values are consistent to within a factor of 1.5 of measured values.

All sensitivities are 5 sigma in one hour of exposure time, in 70%-ile seeing, 50 %-ile transparency, water vapor and background conditions, and at 1.5 airmasses (thus sensitivities can be somewhat better when conditions are very good and/or at low airmasses). Imaging and low resolution spectroscopy use the chop/nod mode with guiding in one beam only and the sensitivity values correspond to 30 minutes of on-source guided integration time, hence 30 minutes on sky, and to about 2.25 hrs of elapsed time, including a set-up time of ~15 minutes and observing overheads. The medium resolution gratings and echelle are used in stare/nod mode, observations in both beams are guided. The sensitivity values correspond to 60 minutes of on-source guided integration time. The total stare/nod elapsed time including overheads is about 1.5 hrs for every 1.0 hrs of exposure time.

In imaging mode the point source sensitivities were measured in apertures giving optimal S/N (i.e. optimal detectivity). For accurate photometry, larger apertures are usually required, so one would expect the S/N to be lower than the ITC values by a factor of ~2.

In spectroscopic mode estimates were made at wavelengths free of strong telluric absorption lines. In the 10um window a 2 pixel-wide slit was employed; in the 20um window a 3 pixel-wide slit was used.

Magnitudes are N or Q for a stellar (Rayleigh-Jeans) continuum. Flux densities are at the wavelength of the filter.

Imaging

Filter R-J Point Source
N or Q (mag)
Point Source
(mJy)
Extended Source
(mJy/arcsec^2)
N' (10.0-12.4um) 11.1 1.1 1.1
Si-1 (7.9um) 9.8 7.2 10.4
Si-2 (8.8um) 10.6 2.9 3.6
Si-3 (9.8um) 10.0 4.0 4.7
Si-4 (10.3um) 10.4 2.5 2.7
Si-5 (11.6um) 10.3 2.1 2.0
Si-6 (12.5um) 9.8 2.9 2.6
Qa (18.1um) 7.3 12 7


Spectroscopy with 2 pixel wide (10um) and 3 pixel wide (20um) slits
(sensitivity per spectral pixel)

Grating - R Wavelength Point Source
(mag)
Point Source
(mJy)
Unresolved Line Flux (W/m2)
LowN - 170 8.5um 8.9 14 2.9E-17
LowN - 210 10.5um 8.5 14 1.9E-17
LowN - 230 12.5um 7.7 20 2.4E-17
LowQ - 100 17.9um 5.1 110 1.5E-16
LowQ - 110 20.1um 4.9 110 1.4E-16
LowQ - 120 21.5um 4.2 150 2.2E-16
LowQ - 130 24.3um 3.5 250 3E-16
MedN1 - 900 9.0um 8.4 21 1.2E-17
MedN1 - 1200 12.0 7.3 35 7E-18
MedN2 - 2700 9.0um 7.7 40 5E-18
MedN2 - 3600 12.0um 6.5 70 5E-18
Ech - ~30000 9.0um 5.7 270 3E-18
Ech - ~30000 12.0um 4.6 400 3.4E-18
Ech - ~20000 18.0um 2.9 700 6E-18



[Science Operations home][Michelle home]


Last updated 3 March 2006; T.R. Geballe and R. S. Fisher